1. Academic Validation
  2. Ultra-performance liquid chromatography electrospray ionization-tandem mass spectrometry method for the simultaneous determination of itraconazole and hydroxy itraconazole in human plasma

Ultra-performance liquid chromatography electrospray ionization-tandem mass spectrometry method for the simultaneous determination of itraconazole and hydroxy itraconazole in human plasma

  • J Pharm Anal. 2014 Oct;4(5):316-324. doi: 10.1016/j.jpha.2013.09.005.
Ashish Dwivedi 1 Bhupinder Singh 1 Sandeep Sharma 1 R S Lokhandae 2 Naveen Dubey 1
Affiliations

Affiliations

  • 1 Bioanalytical Department, Jubilant Life Sciences, C-46, Sector 62, Noida 201307, India.
  • 2 Jaipur National University, Jaipur, India.
Abstract

A highly sensitive, selective, and precise ultra-performance liquid chromatography tandem mass spectrometry method was developed and validated for simultaneous quantification of itraconazole and hydroxy itraconazole in human plasma by a single liquid-liquid extraction step. The precursor to product ion transitions of m/z 705.3/392.3, m/z 721.2/408.3 and m/z 708.2/435.4 were used to detect and quantify itraconazole, hydroxy itraconazole and itraconazole-d3 respectively. The lower limit of quantitation was found to be 0.500 ng/mL for itraconazole and 1.00 ng/mL for hydroxy itraconazole. The mean recoveries for itraconazole and hydroxy itraconazole were found to be 100.045% and 100.021%, respectively. This developed method with a chromatographic run time of 2.0 min was successfully applied to a bioequivalence study of 100 mg itraconazole capsule.

Keywords

Human plasma; Hydroxy itraconazole; Itraconazole; Simultaneous analysis; Ultra-performance liquid chromatography.

Figures
Products