1. Academic Validation
  2. Quantitative analysis by high-performance liquid chromatography atmospheric pressure chemical ionization mass spectrometry: The determination of the renin inhibitor CP-80,794 in human serum

Quantitative analysis by high-performance liquid chromatography atmospheric pressure chemical ionization mass spectrometry: The determination of the renin inhibitor CP-80,794 in human serum

  • J Am Soc Mass Spectrom. 1991 Apr;2(2):164-7. doi: 10.1016/1044-0305(91)80010-5.
H Fouda 1 M Nocerini R Schneider C Gedutis
Affiliations

Affiliation

  • 1 Drug Metabolism Department, Central Research Division, Pitzer, Inc., 06340, Groton, CT.
Abstract

A specific method was developed for the quantitative determination of the Renin Inhibitor CP-BO,794. Serum extracts containing the drug and an internal standard were injected into a standard reverse-phase high-performance liquid chromatography (HPLC) column. The mobile phase, methanol/water (8/2), flowed at 1 ml/min through the column and then via a heated nebulizer interface into a corona discharge atmospheric pressure chemical ionization source. The assay minimum limit of quantification was 50 pg/mL. It exhibits satisfactory accuracy and precision over the range 50 pg/ml, to 10 ng/mL. A minor modification of the HPLC mobile phase was necessary to attain extremely low detection limits. The addition of a structural analogue contributed to enhancing the precision of the assay.