Organs-on-chips with integrated electrodes for trans-epithelial electrical resistance (TEER) measurements of human epithelial barrier function
- Lab Chip. 2017 Jun 27;17(13):2264-2271. doi: 10.1039/c7lc00155j.
- 1. Wyss Institute for Biologically Inspired Engineering, Harvard University, CLSB5, 3 Blackfan Circle, Boston, MA 02115, USA. [email protected].
Trans-epithelial electrical resistance (TEER) is broadly used as an experimental readout and a quality control assay for measuring the integrity of epithelial monolayers cultured under static conditions in vitro, however, there is no standard methodology for its application to microfluidic organ-on-a-chip (organ chip) cultures. Here, we describe a new microfluidic organ chip design that contains embedded electrodes, and we demonstrate its utility for assessing formation and disruption of barrier function both within a human lung airway chip lined by a fully differentiated mucociliary human airway epithelium and in a human gut chip lined by intestinal epithelial cells. These chips with integrated electrodes enable real-time, non-invasive monitoring of TEER and can be applied to measure barrier function in virtually any type of cultured cell.