Development of a highly sensitive chemiluminescent assay for hydrogen peroxide under neutral conditions using acridinium ester and its application to an enzyme immunoassay
- Luminescence. 2014 Jun;29(4):374-7. doi: 10.1002/bio.2555.
- 1. School of Pharmacy, Showa University, Shinagawa-ku, Tokyo, 142-8555, Japan.
We developed a highly sensitive chemiluminescent (CL) assay for hydrogen peroxide using 10-methyl-9-(phenoxycarbonyl) acridinium fluorosulfonate (PMAC) that produced chemiluminescence under neutral conditions and applied it to an enzyme Immunoassay (EIA). One picomole of hydrogen peroxide could be detected using the optimized PMAC-CL method and 6.2 × 10(-20) mol β-D-galactosidase (β-gal) could be detected by combining an indoxyl derivative substrate and the proposed PMAC-CL method. This highly sensitive CL β-gal assay was applied to an EIA for thyroid-stimulating hormone (TSH) using β-gal as a label enzyme; 0.02-100.0 μU/mL TSH in human serum could be assayed directly and with high reproducibility.
-
Cat. No.Product NameDescriptionTargetResearch Area
-
target: ROS KinaseResearch Areas: Others