The use of chlorophenol red for the selective determination of chlorine dioxide in drinking water

  • Talanta. 1996 Jan;43(1):103-8. doi: 10.1016/0039-9140(95)01721-6.
D L Sweetin ,  E Sullivan ,  G Gordon
Abstract

In recent years, the use of chlorine dioxide as an alternative disinfectant for drinking water has become increasingly attractive. As a result, an accurate method for the determination of mg l(-1) concentrations of chlorine dioxide is needed. Improvements to chlorophenol red (CPR) spectrophotometry result in a selective method for ClO(2) with few interferences. CPR selectively reacts with 0.1-1.9 mg l(-1) ClO(2) at pH 7, yielding a linear response (0.9994) with a limit of detection of 0.12 mg l(-1) ClO(2). Several species, ClO(2)(-), ClO(3)(-), NH(2)Cl, and free available chlorine (FAC), were studied as potential interferents using this method. There was found to be less than 2% interference due to 1.38 mg l(-1) ClO(2)(-), 9.87 mg l(-1) ClO(3)(-1), and 5.31 mg l(-1) NH(2)Cl. The interference from up to 1.19 mg l(-1) FAC was 3.7% and could be further reduced by the addition of oxalic acid, sodium cyclamate or thioacetamide.