PE-Cy7
Based on 1 Customer Validation
PE-Cy7 is a tandem dye conjugate based on fluorescence resonance energy transfer (FRET). Composed of phycoerythrin (PE) as the donor and Cy7 as the acceptor fluorescent dye, PE-Cy7 is suitable for multicolor flow cytometry, laser scanning cytometry and immunofluorescence applications. PE-Cy7 can be conjugated to antibodies or used in the biotin-streptavidin system to label cell surface antigens in immunophenotyping and multiparametric cell analysis. PE-Cy7 undergoes photon-induced oxidative degradation, whose rate depends on light intensity and storage conditions, leading to reduced FRET efficiency and increased PE leakage. PE-Cy7 can be excited at wavelengths of 488 nm, 532 nm or ~760 nm, and its emission signal can be detected using 780/60 nm, 810/90 nm or 785/50 nm bandpass filters, with a Stokes shift of 300 nm.
Nur für Forschungszwecke. Wir verkaufen nicht an Patienten.
-
Speicherung:
4°C, protect from light
* In solvent : -80°C, 6 months; -20°C, 1 month (protect from light)
Biologische Aktivität
Beschreibung
In Vitro
Operating Instructions (The following is a recommended procedure for reference only, and adjustments are required based on specific requirements)
1. Stock Solution Preparation
1.1 Solvents:
1.1.1 Storage buffer containing 0.016 M phosphate buffer (pH 7.4), 0.02% sodium azide and 0.2% bovine serum albumin.
1.1.2 Anhydrous DMSO.
1.2 Concentration Recommendations:
1.2.1 The concentration of bis-Cy7 stock solution is 10 mg/mL (10 mM).
1.2.2 Note: No stock solution concentration is specified for PE-Cy7 supplied as a conjugated antibody/streptavidin conjugate.
2. Working Solution Preparation
2.1 Diluent:
2.1.1 Dilute PE-Cy7-conjugated streptavidin or antibody at a ratio of 1:10 before use.
2.1.2 Note: PE-Cy7 supplied as a conjugated antibody/streptavidin conjugate requires no dilution and can be used directly.
2.2 Working Concentration: Optimize the concentration via titration to achieve saturated staining.
2.3 Note: Adjust the working solution concentration as needed; prepare and use it immediately.
3. Staining Procedure
3.1 Sample Types: Suitable for suspension cells (human peripheral blood mononuclear cell fractions, human peripheral blood leukocytes, peripheral blood mononuclear cells (PBMC))[1][2][3]:
3.1.1 If whole blood leukocytes are used, process EDTA-anticoagulated blood samples.
3.1.2 Use at least 106 cells per stained sample.
3.1.3 Wash the cells twice before staining.
3.2 Incubation Conditions:
3.2.1 Direct staining: Incubate cells with PE-Cy7-conjugated antibody for 15 min at room temperature in the dark, or for 15 min at 0℃ (ice bath); optimize conditions as needed.
3.2.2 Indirect staining: Aliquot blood samples and incubate with biotinylated primary antibody for 15 min at room temperature, then add PE-Cy7-conjugated streptavidin diluted at 1:10, and continue incubation for 15 min at room temperature.
3.3 Washing Steps:
3.3.1 After staining is completed, lyse red blood cells, then wash the samples with phosphate buffer.
3.3.2 Wash three times with staining medium (RPMI medium without biotin and flavins, supplemented with 4% fetal bovine serum and 0.02% sodium azide).
4. Control Setup
4.1 Set up unstained control, single-positive control, and Full-Minus-One control samples.
4.2 Positive controls are used to validate the experimental system; blank controls are used to eliminate reagent fluorescence interference.
5. Detection and Analysis
5.1 Instrument Types: Flow cytometer, laser scanning cytometer.
5.2 Excitation/Emission Wavelengths:
5.2.1 Excitation light: 532 nm laser, 488 nm argon ion laser.
5.2.2 Emission light: Detect via a 780/60 bandpass filter, detect via an 810/90 nm bandpass filter, or collect emission light >780 nm via a 50 nm bandpass filter with a central wavelength of 785 nm.
5.3 Result Analysis:
5.3.1 Changes in Fluorescence Intensity:
5.3.1.1 Degradation of PE-Cy7 leads to increased PE leakage (the ratio of donor dye emission to acceptor dye emission increases).
5.3.1.2 The PE-Cy7 fluorescence intensity of target-positive cell populations (e.g., CD3-positive lymphocytes) is significantly higher than that of target-negative cell populations.
5.3.1.3 Before reaching the threshold, Cy7 fluorescence intensity increases with the rise of the molar ratio of Cy7 to PE; beyond the threshold, self-quenching causes a decrease in Cy7 fluorescence intensity; meanwhile, due to enhanced resonance energy transfer, PE fluorescence intensity decreases with the rise of the molar ratio of Cy7 to PE.
5.3.2 Fluorescence Localization: Targets to cell surface antigens via conjugated antibodies.
5.3.3 Other Parameters:
5.3.3.1 The resolution parameter of PE-Cy7 for distinguishing positive and negative cell populations is comparable to that of FITC and PE-Cy5.
5.3.3.2 Significant fluorescence spillover occurs into PE and PE-Cy5 channels; minimal spillover occurs into APC and APC-Cy7 channels.
5.3.3.3 Emits far-red fluorescence.
5.3.3.4 Instrument Settings (for laser scanning cytometer): Set the photomultiplier tube gain to 255, the offset to 2075, and the photomultiplier tube voltage to 55% of the maximum voltage.
MedChemExpress (MCE) has not independently confirmed the accuracy of these methods. They are for reference only. Further protocols information, click here.
Chemical Information
-
Appearance Lyophilized Powder
-
Color White to off-white
-
SMILES
[PE-Cy7]
-
Versand
Room temperature in continental US; may vary elsewhere.
-
Speicherung
4°C, protect from light
* In solvent : -80°C, 6 months; -20°C, 1 month (protect from light)
Protokoll
-
Flow Cytometry
Flow cytometry (FC) is a technique for high-speed, step-by-step quantitative analysis and sorting of single cells or other biological particles in a suspension by detecting labeled fluorescent signals.
-
Flow cytometric DNA-content cell-cycle staining
Flow cytometric DNA-content cell-cycle staining measures the fluorescence intensity of DNA-bound fluorochromes in single cells or nuclei to estimate DNA content distributions, allowing assignment of populations to G0/G1, S, and G2/M phases by DNA histogram deconvolution. Propidium iodide (PI) intercalates into DNA, and PI fluorescence is proportional to cellular DNA content when staining is performed under conditions that make DNA accessible and minimize non-DNA signal. Cells with G2/M DNA content are expected to show approximately twice the fluorescence intensity of G0/G1 cells, while S-phase cells occupy intermediate fluorescence values. PI-based DNA-content analysis can also detect cells with fractional DNA content, often reported as sub-G1, when DNA fragmentation and extraction during staining reduce retained DNA signal in apoptotic cells. DAPI is an alternative DNA fluorochrome for univariate DNA-content analysis, while bivariate approaches combining DNA content with proliferation
-
Protocol for Phospho-flow cytometry
Phospho-flow cytometry detects intracellular phosphorylated signaling proteins in single cells using phospho-specific antibodies after rapid fixation and permeabilization; the fluorescence intensity reflects phosphorylation state and therefore kinase-pathway activation, inhibition, or drug response in defined cell subsets. Unlike Western blot, phospho-flow preserves single-cell resolution and can measure signaling heterogeneity in cancer cells, primary immune cells, dissociated mouse tumors, macrophages, organoid-derived cells, and drug-screening samples when validated antibodies and fixation/permeabilization conditions are used.
-
Antibody-based immunofluorescence/immunocytochemistry staining
Antibody-based immunofluorescence/immunocytochemistry detects the cellular or subcellular localization of a target antigen by binding a primary antibody to the target and detecting that antibody directly with a fluorophore-conjugated primary antibody or indirectly with a fluorophore-conjugated secondary antibody. Indirect immunofluorescence can amplify signal because multiple secondary antibodies can bind one primary antibody. The assay readout is fluorescence intensity and localization measured by fluorescence or confocal microscopy, and the result reflects antigen distribution only when the antibody has been validated for the target, sample type, fixation condition, and imaging workflow. Antibody specificity must not be assumed from catalog information alone, and appropriate validation or control experiments are required for serious interpretation.
- Immunocytochemistry/Immunofluorescence
Reinheit & Dokumentation
-
Data Sheet (281 KB)
-
SDS (252 KB)
- English - EN (252 KB)
- Français - FR (252 KB)
- Deutsch - DE (252 KB)
- Norwegian - NO (252 KB)
- Español - ES (252 KB)
- Swedish - SV (252 KB)
- Italian - IT (252 KB)
- Korean - KR (252 KB)
- Portuguese - PT (252 KB)
-
Handling Instructions (2659 KB)
Verweise
Calculators
Konzentration (Stammlösung) × Volumen (Stammlösung) = Konzentration (Ziellösung) × Volumen (Ziellösung)